Application
For gaseous, dry media which do not attack copper alloys.
! For measuring gas or vapour, observe the table “Selection Criteria as per EN 837-2” (see appendix)!
Accuracy class (EN 837-3/6)1,6
Ranges (EN 837-3/5)See ordering table
Application area
Static load:Full scale value
Dynamic load:0.9 x full scale value
Overload safety:1.3 x full scale value
Operating temperature range
Medium:Max. 60 °C
Ambient:-20/+60 °C
Temperature performanceIndication error when the temperature of the measuring system deviates from the normal temperature of 20 °C:
at rising temperature approx. ± 0.6 %/10 K
falling temperature approx. ± 0.6 %/10 K
of full scale value
Degree of protection
NG 100:IP 54 (EN 60529)
Measuring elementCapsule element, CuBe alloy
Zero correctionFrom the front
DialAluminium, white
Scaling black
HousingStainless steel 304
Bayonet type bezelStainless steel 304
Window
D401:Instrument glass
Options
- Overpressure and underpressure safety 10 x FSD
- Back flange
- Panel mounting bezel, with window, plastic
- 3-hole fixing, panel mounting bezel with window, plastic
- Damping screw
- Reference pointer
- Special scales
- Other process connections