SLK

    Categories Liquid & Gas Analysis, Conductivity Transmitters

    Brand : Seli

    • Description
    • Attachment

    Features

    • material:
      probe tip: 316L/1.4404/1.45771
      insulator: Peek (FDA)
      housing parts: 1.4305/1.4301
      Nozzle G1/2” Sw22: 1.4305/1.4301
    • Digital PNP output or analogue output
    • settings via PC software or Y-programming adapter
    • weld-in sleeve system with modular process connections
    • aseptic measuring point
    • electrodes can be shortened independently
    • medium-affected components are made of Peek
    • FDA, EHEDG conform

    Technical specifications

    • supply voltage  Ub = 24 V +/- 20% (18…32 VDC)
    • measuring range: 0 µS/cm – 15000 µS/cm
    • output signal; PNP; 18-32 V; permissible load 0 OHM at 24 V; 35 mA or alternatively analogue 4-20 mA; burden <= 680 Ohm
    • response time <0.5s
    • ambient temperature: -10…+60ºC
    • process temperature: 0…+100ºC
    • thermowell version: 0…+125ºC
    • storage temperature -20…+70ºC
    • CIP- / SIP cleaning: 0…+150ºC (30 min)
    • operating pressure: max. 10 bar
    • accuracy: 5% of measurement value
    • integrated temperature compensation
    • protection class IP 68
    • tightening torque 5…20 Nm

    Data Sheet

    Download

    Manual

    Download