DOUBLE CHANNEL ANALYZERS SERIES 46

    Categories Liquid & Gas Analysis, pH Transmitter & Controller

    • Graphic display and keypad 128 x 128 pixel monochrome display with graphic icons to show digital output status, washing cycle and alarms with three-colour backlight. Simultaneous flashing values for the measurement (numeric + paragraph) and temperature readings.
    • Five control keys for instrument calibration and configuration
    • Data Logger with 260,000 records downloadable by accessing the internal web server with Wi-Fi connection
    • Enclosure Box
      • Wall mounting ABS plastic material IP65 full box (144x144x122)
      • Panel mounting ABS IP65 front panel only (96x96x42)
    • All outputs Relay, SSR and Output mA are configurable with primary and secondary measure
    • Description
    • Attachment

      MEASURES

    • pH / ORP (redox)
    • Conductivity
    • Chlorine
    • Chlorine dioxide
    • Ozone
    • Hydrogen peroxide
    • Peracetic acid
    • Chlorites Bromine
    • Dissolved oxygen
    • Turbidity
    • Suspended solids

    TECHNICAL DATA

    Measure Ranges pH: 0.00 – 14.00
    ORP: -2000mV – 2000mV

    Conductivity:
    with C 0.01 cm⁻¹ / K100 sensor: 0.005µS/cm to 200µS/cm
    with C 0.10 cm⁻¹ / K10 sensor: 0.05µS/cm to 2mS/cm
    with C 0.20 cm⁻¹ / K5 sensor: 0.1µS/cm to 4mS/cm
    with C 1.00 cm⁻¹ / K1 sensor: 0.5µS/cm to 20mS/cm
    with C 10.0 cm⁻¹ / K0.1 sensor: 5µS/cm to 200mS/cm
    with C 20.0 cm⁻¹ / K0.5 sensor: 10µS/cm to 400mS/cm

    Input mA: Depending by the type of sensor
    Sensor power supply 4/20mA 2 or 4 wires (*) 18Vdc ±5%, max 30mA
    Measure Accuracy pH: ±0.01 pH
    ORP: ±1 mV
    Conductivity: ±2% on the measuring point
    Input mA: ±0.2%
    Temperature Input Pt100 / Pt1000 Automatic Recognition
    Display Graphic TFT colour LCD
    Resolution 128 × 128 pixels or 240 × 128 pixels
    Languages Italian, English, French, German, Spanish
    Keypad 5 bubble-keys
    Data Logger 260000 records, minimum interval 1 min, data download via Wi-Fi
    Analogue outputs Two (2) programmable: 4…20 mA
    Galvanic separation
    Maximum load 800 Ohm
    NAMUR: OFF 3.6 mA, 22 mA
    Digital Outputs Four (4): Two set points and two alarms
    Set point (2) Set points: RL1 and RL2 2-SPST mechanical 250 VAC / 5A, 30 VDC / 3A
    RL1 and RL2 relay configurable Load activation
    RL3 and RL4 relay configurable Load activation, Probe washing, Alarm repetition
    Setting Cycle time and Delay time
    Frequency output Two (2) SSR1-SSR2 SPST 60 VAC max 100 mA
    Bidirectional, NPN, PNP
    Resistance in ON state 5 ohm @100mA
    Frequency range 0 to 400 pulses/min
    Pulse Duration 100 ms
    Digital Input REED digital input for dry contact 5 VDC, max 6 mA
    Communication port RS485 modbus server ASCII / RTU standard protocol
    Output 24Vdc 24Vdc ±5% max 250mA
    Operating conditions Operating temperature: -10 – 50°C
    Storage: -25 – 65°C
    Humidity: 10-95% (non-condensing)
    Power supply / Electrical protections Universal power supply 100-240 Vac 50/60 Hz
    Low voltage power supply 24-48 Vdc or 24 Vac CE compliant
    Transformer isolation 4KV
    Absorbing average 3.6A
    Electrical Protection: EMI / RFI CEI-EN55011 – 05/99
    According to UL, not connect to relay outputs exceeding 115V
    Mechanical protections Wall mounting: IP65
    Panel mounting: IP65 Front, IP20 Back

    Data sheet

    Download