MULTI CHANNEL ANALYZERS SERIES 50

    Categories Liquid & Gas Analysis, pH Transmitter & Controller

    • Available in three configurations: Up to two, four and eight simultaneous measurements, freely selectable
    • Connectable to the entire range of Chemitec digital sensors, plus an universal input for 4-20mA type sensors
    • Simultaneous display of digital measurements via 480x272 display with color variation depending on alarm events. Display of measurement trends in graphic form
    • Internal data logger with 250,000-record capacity, data download via USB or RS485
    • Programmable analogue outputs for repeating measurements, PID control and temperature
    • Digital output relays for set point adjustment
    • Analogue input for perturbative functions or engineered display of additional measuring
    • Digital input for disabling of dosage
    • Description
    • Attachment

      MEASURES

    • pH/ORP
    • Dissolved oxygen
    • Conductivity
    • Turbidity
    • Suspended solids
    • Chlorine
    • Chlorine dioxide
    • Ozone
    • Chlorites
    • Hydrogen peroxide
    • Peracetic acid
    • Nitrates (ISE)
    • Nitrates (UV)
    • Organic substances (UV)
    • Colour (UV)
    • Chlorophyll
    • PAH/OIL (UV-Fluorescence)

    TECHNICAL DATA

    Measure ranges Depending by the connected sensor
    Temperature Input Pt100 / Pt1000 Automatic Recognition
    Visualization LCD Graphic color display 480 (R,G,B.) (W) × 272 (H)
    Programming 5 bubble-Keys keyboard
    Data Logger Flash 32Mbit Memory (near to 250000 records)
    Languages Italian, English, French, German, Spanish, Polish
    4 Analogue Outputs 0 / 4.00 – 20.00 mA
    Galvanic separation: 1KV Optoisolator
    Maximum load 500 Ohm
    Second Alarm output: NAMUR 2.4 mA (with 4/20mA Range)
    Possibility of PID adjustment
    4 command digital outputs
    Max 2 for each measure
    Set Point ON–OFF: working range setting (Hysteresis / direction) and pause/working time setting: 000 – 9999 Seconds
    PID Regulation (only on Set point 1): PWM
    Switching relay
    Maximum resistive load 3A at 230Vac
    2 digital outputs for alarm / wash Reporting: Instrumental anomalies, minimum, maximum, set point’s delay, permanence time (live check) warning on probe calibrations
    Display data and time of last calibration
    Calibration warning on screen
    Delay time: 00 – 99:99mm:ss minimum steps of 15 sec
    Permanence time: 00:00 – 99:99 hh:mm
    Set Point disabling in case of alarm: Enable / Disable
    Wash: Programming of time leg
    Frequency: 00:00 – 24:00 hh:mm minimum time leg: 15 min
    During washing phase all digital and analogue outputs are freezed
    Switching relay
    Maximum resistive load 3A at 230Vac
    Powered digital input To disable dosages or activate washing cycle
    Possibility to link with a 3 wires inductive sensor
    RS485 Serial output For set-up and real-time data acquisition from remote or for stored data download (using a dedicate-SW)
    MODBUS RTU communication protocol
    1200 – 38400 Baud Rate programmable speed
    Analogue Input 4-20mA analog input with value engineering
    12Vdc power supply for 2 and 4 wire sensors
    Operating conditions Operating temperature -20 – 65°C
    Storage and transport -25 – 65°C
    Humidity 10-95% (non-condensing)
    Power supply / Electrical protections Power supply 100–240Vac/dc 50–60 Hz – 24Vac/dc
    Transformer isolation 4KV
    Absorbing average < 7W
    Electrical Protection: EMI / RFI CEI-EN55011 – 05/99
    According to UL, not connect to relay outputs exceeding 115V

    Data sheet

    Download