S423 C OPT

    Categories Liquid & Gas Analysis, Dissolved Oxygen

    • Reliable concentration measurement using optical measuring process
    • Dynamic luminescence measuring method
    • AISI 316 or Black rigid PVC sensor body
    • Interchangeable cap for luminophore's replacement
    • No mechanically moving parts
    • Immediate installation and easy maintenance
    • Ability to se salinity and barometric pressure for oxygren value

      APPLICATIONS

    • Measure of oxygen in wastewater
    • Measure of oxygen in primary, industrial, recirculating water
    • Description
    • Attachment

    TECHNICAL DATA

    Materials AISI 316 and PVC body • Special glass optics • NBR and silicon O-rings
    Thread 3/4” BSP
    Measuring ranges 0.00 mg/L to 20.00 mg/L
    Measuring method Optical measure by luminescence
    Calibration method Calibration on air: 1-point calibration, water saturated with air at 100%
    Calibration through sample: comparison with a standard tool
    Calibration on 0% solution: 1-point calibration, specific solution at 0% oxygen saturation
    Resolution 0.01 mg/L
    Accuracy ±0.2 mg/L when < 5 mg/L
    ±0.3 mg/L when > 5 mg/L
    Repeatability ±0.1 mg/L
    Response T90 < 60s
    Maximum refreshing time < 1 second
    Working temperature 0 – 50 °C (Accuracy ±0.2°C)
    Max working pressure 5 bar
    Max absorption 2W
    Mechanical protection IP68 Sensor + cable
    Cable 10m integral
    Power supply 12…24Vdc
    Signal interface RS-485 Modbus RTU Protocol (4–20mA optional)
    Water move No necessary move
    Compensation of temperature Via internal NTC
    Luminophore diameter 10 mm
    Connector IP67 Yes

    Data sheet

    Download